© Adam Equipment Company 2020
22
•
Remove the sample weight. Then any other weight placed on the scale will be
displayed as a percentage of the original sample. For example, if 3500g is placed
on the scale and percent weighing is selected, the display will show 100.00%.
Remove the 3500g weight and place a 3000g weight. The display will show 85.7%
as 3000g is 85.7% of 3500g.
•
The number of decimal points will depend on the weight used. A smaller weight will
show only “100%” while a larger weight might show “100.00%”.
•
If the scale was showing zero weight before entering this function, then the user
must manually enter the weight to be set as 100%. When
“F4 PCT”
is displayed,
enter the weight to be used for the 100% reference, then press [Tare] to accept the
reference weight. The display will show “0.00 %”.
•
If the scale shows
“x x . x x %”,
which is the last weight used as a reference, press
[CE] to clear and use the numeric keypad to enter a new value. Press [Tare] to
accept the new reference weight.
•
The weight entered must be greater than 50 scale divisions.
•
Press [Unit] to return to normal weighing.
NOTE: The display may jump by large numbers unexpectedly if small weights are used to
set as 100% reference. The scale checks if the weight is too small and will show Error 8.