© Adam Equipment Company 2020
20
10.6 LIMITS STORING AND RECALLING
The scale can store up to 10 sets of high and low limits in memory along with the weighing
units in use (including pcs for parts counting and % for percent weighing) as well as
settings for the beeper.
During Check weighing the current limits can be stored or previously stored units can be
recalled
.
Press the [
Lim] key. If you are already in the
check weighing mode the display will ask if you
wish to store the current limits by showing
“StOrE”
or recall another set of limits by showing
“
rECALL
”. The [
Lim] key can be used to toggle
between
“StOrE
” and “
rECALL
”.
If you want to store the limits, when
“StOrE
” is
displayed press the [Tare] key. The display
shows “
St
”. Enter a number corresponding to
the desired memory location (0 to 9). “
St X
” will
be displayed for 2 seconds indicating the
location X where the current limits, weighing
units and settings for the beeper bar graph are
stored. The scale will continue to work with the
current settings as active.
If you wish to recall any of the pre-stored limits,
press [Tare] when
“rECALL”
is displayed. The
display shows “
rEC
”. Enter the number
corresponding to the desired memory location (0
to 9) to be recalled. “
rEC X
” will be displayed for 2
seconds indicating the values stored in the
location “X” is being recalled. The scale will
change to the recalled limits, weighing units and
settings for the beeper bar graph.