Fault detected
Possible causes
Solutions
Device does not operate
(LED indicator off)
Power cord not plugged in properly.
- Check the power socket.
- Return to SATELEC after-sales.
Power power switch in position O.
- Place the power switch in position I.
Power receptacle fuse(s) blown.
Replace the power receptacle fuses.
Internal fuse blown.
Return to SATELEC after-sales.
No spray
Irrigation solution tank empty.
Fill the irrigation solution tank.
Faulty pump.
Replace the pump.
Irrigation deactivated
Return to SATELEC after-sales
Blocked irrigation circuit
Return to SATELEC after-sales.
Blocked tip or file.
Unblock the tip or file.
Wrong choice of tip.
Check the tip.
Power fault/
low vibrations
Tip worn or distorted.
Replace the tip.
Incorrect power adjustment.
Refer to the TipBook for the necessary
information.
Incorrect use: incorrect angle of attack or
inadequate pressure on the tooth..
Refer to the TipBook for the necessary
information.
Liquid or moisture present between
handpiece and cord.
Dry the electrical contacts thoroughly.
No ultrasonic vibration
Incorrect tip tightening.
Tighten the tip with the wrench.
Faulty connector contact.
Clean the connector contacts.
Handpiece cord wire cut.
Return to SATELEC after-sales to replace
the cord.
Leak between the
handpiece cord and the
connector on the device
Worn handpiece 1.15x1 seal.
Replace the seal (kit F12304).
Leak of liquid under the
device
Tube rupture.
Return to SATELEC after-sales.
Leak of liquid at the tank
connector level.
Worn seals.
- Replace the two seals (E15121).
- Return to SATELEC after-sales.
Summary of Contents for SATELEC P5 Newtron XS
Page 1: ......
Page 2: ...2 ...
Page 3: ...English 4 Français 22 Español 40 Deutsch 58 Italiano 76 Nederlands 94 3 ...
Page 21: ...21 ...
Page 39: ...39 ...
Page 57: ...57 ...
Page 75: ...75 ...
Page 93: ...93 ...
Page 111: ...111 ...
Page 114: ...114 19 20 9 8 13 14 16 17 18 15 12 1 1 1 0 3 4 5 1 Fig Abb 1 2 3 1 4 Fig Abb 2 ...
Page 115: ...115 3 2 1 4 Fig Abb 3 1 Fig Abb 4 ...