P/N 800-1090-104 Rev L102012
Page 26 of 44
Measurement Accumulator
This Function allows multiple measurements to be
made, and the sum displayed on the Readout.
S
ET
P
ROGRAMMING
P
ARAMETER
pr24
=
3
to enable this function.
P
ROGRAMMING
P
ARAMETER
is used to configure this function.
F3 & F4
are used to execute this function.
1. Press the
F4
key to view any previous accumulator value.
The LCD will display
Accu
for 1 second followed by the total
accumulated measurements. The display format for the accumulated total
is controlled by both the currently selected measurement mode (
UNITS
key) and the Programming Parameter
a.
If
Pr25
is set to 0, and the Readout is in inch mode, (decimal or
fractions), the accumulated total is displayed in Decimal Inches.
b.
If
Pr25
is set to 0, and the Readout is in metric mode, (mm or cm),
the accumulated total is displayed in mm.
c.
If
Pr25
is set to any non-zero value, and the Readout is in inch
modes, (decimal or fractions), the accumulated total is displayed
as feet and inches. (fff ii) where f=feet and i = inches
d.
If
Pr25
is set to any non-zero value, and the Readout is in metric
mode, (mm or cm), the accumulated total is displayed as meters
and millimeters. (MM mmm) where M=meters and m=millimeters.
2. Press the
F3
key to clear the accumulator.
The LCD will display
CLr
for 1 second. The DRO will return to the normal
measuring mode.
3. To add a measurement to the accumulator, make the measurement and
then press the
F3
key. The LCD momentarily displays
Add
to indicate
that the measurement has been added to the accumulated total.
Continue to add measurements by pressing the F3 key as each
measurement is made.
4. To view the accumulated total, press the
F4
key.
The following options are now available:
a. Press
F3
to clear the total and return to the measuring mode.
b. Press
F4
(DOES NOT clear the total) to return to the measuring mode.
c. Press
SEND
to transmit the accumulated total to the output port.
NOTE
: This Function maintains a running sum of the measurements taken.
Individual measurements in a series cannot be edited or deleted. If an error is
made and an incorrect measurement is taken, the entire summation must be
cleared and repeated. No rounding of inches or millimeters occurs when
configured for feet/inches or meters/millimeters.