Display Shows Possible Cause(s)
Possible Solution(s)
Test Strips Expiring
j
The test strips expire at the
end of the current month.
j
This message may appear when using a white code
key in the meter. It means the test strips expire at
the end of the current month. At the end of the
month, discard the white code key and any
remaining test strips. Insert a black code key. Make
sure the time and date in the meter are correct. If
you need a black code key, please contact
ACCU-CHEK
Customer Care Service Center at
1-800-688-4578.
j
Ensure the time and date are correct and adjust, if
necessary. See Chapter 5, Changing Meter
Settings.
Time Mismatch
j
The time or date on the meter
is more than five minutes
different than the time or date
on the pump.
j
The meter time and date were changed to match
the pump time. Ensure the time and date are
correct. If the time or date are not correct, change
the time and date on the pump.
51638_00048010155_C.indb 262
6/11/13 3:58 PM
Summary of Contents for Aviva
Page 2: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...
Page 12: ...6 Glossary 293 Index 301 51638_00048010155_C indb 6 6 11 13 3 57 PM ...
Page 44: ...38 51638_00048010155_C indb 38 6 11 13 3 57 PM ...
Page 142: ...136 51638_00048010155_C indb 136 6 11 13 3 58 PM ...
Page 250: ...244 51638_00048010155_C indb 244 6 11 13 3 58 PM ...
Page 270: ...264 51638_00048010155_C indb 264 6 11 13 3 58 PM ...
Page 313: ...307 51638_00048010155_C indb 307 6 11 13 3 58 PM ...
Page 314: ...308 51638_00048010155_C indb 308 6 11 13 3 58 PM ...
Page 315: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...