Display Shows Possible Cause(s)
Possible Solution(s)
Communication
Warning
j
Communication of bolus data
from the pump was not
successful. Therefore, pump
data are not available and the
bolus data may not be
accurate.
j
Communication of bolus data
from the pump was not
successful. Therefore, pump
data are not available and the
active insulin amount may not
be accurate.
j
Ensure the meter and pump have
Bluetooth
wireless
technology turned on.
j
Ensure the meter and pump are within
communication range.
j
If the low battery icon is displayed on the meter, the
meter cannot communicate with the pump. Replace
the batteries.
51638_00048010155_C.indb 252
6/11/13 3:58 PM