Display Shows Possible Cause(s)
Possible Solution(s)
E-53 Bad Test
Error
j
A meter or test strip error has
occurred.
j
Discard the test strip and repeat the test.
j
Your blood glucose may be
extremely high.
j
If this matches how you feel, contact your
healthcare professional immediately. If it does not
match how you feel, repeat the blood glucose test
and see Chapter 2, Testing Your Blood Glucose. If
this still does not match the way you feel, run a
control test with your control solution and a new
test strip. If the control result is within the
acceptable range, review the proper testing
procedure and repeat your blood glucose test with a
new test strip. If the E-53 error still appears for your
blood glucose test, your blood glucose test result
may be extremely high and above the system’s
reading range.
Contact your healthcare
professional immediately.
If the control test result
is not within the acceptable range, see Chapter 3,
Control Testing.
51638_00048010155_C.indb 255
6/11/13 3:58 PM
Summary of Contents for Aviva
Page 2: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...
Page 12: ...6 Glossary 293 Index 301 51638_00048010155_C indb 6 6 11 13 3 57 PM ...
Page 44: ...38 51638_00048010155_C indb 38 6 11 13 3 57 PM ...
Page 142: ...136 51638_00048010155_C indb 136 6 11 13 3 58 PM ...
Page 250: ...244 51638_00048010155_C indb 244 6 11 13 3 58 PM ...
Page 270: ...264 51638_00048010155_C indb 264 6 11 13 3 58 PM ...
Page 313: ...307 51638_00048010155_C indb 307 6 11 13 3 58 PM ...
Page 314: ...308 51638_00048010155_C indb 308 6 11 13 3 58 PM ...
Page 315: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...