253
9
Display Shows Possible Cause(s)
Possible Solution(s)
Connection Lost
j
The connection was lost
between the meter and the
pump during the pairing
process. Therefore, the
attempt to pair the meter and
pump was unsuccessful.
j
Restart the pairing process. See the Advanced
Owner’s Booklet.
j
Ensure the meter and pump have
Bluetooth
wireless
technology
turned on.
j
If the low battery icon is displayed on the meter, the
meter cannot communicate with the pump. Replace
the batteries.
j
Pump is out of range of the
meter.
j
Ensure the meter and pump are within
communication range.
Diary Entry Used
For Advice
j
The diary entry selected has
been used for bolus advice
and modifications are not
allowed.
j
Carefully confirm all information involving bolus
advice. You cannot modify bolus advice-related
entries on the meter.
Diary Results
Expired
j
Bolus advice is no longer
available for this result.
j
Review data that was saved in My Data. Then,
decide to modify data, add data, or begin a new
bolus advice session.
51638_00048010155_C.indb 253
6/11/13 3:58 PM
Summary of Contents for Aviva
Page 2: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...
Page 12: ...6 Glossary 293 Index 301 51638_00048010155_C indb 6 6 11 13 3 57 PM ...
Page 44: ...38 51638_00048010155_C indb 38 6 11 13 3 57 PM ...
Page 142: ...136 51638_00048010155_C indb 136 6 11 13 3 58 PM ...
Page 250: ...244 51638_00048010155_C indb 244 6 11 13 3 58 PM ...
Page 270: ...264 51638_00048010155_C indb 264 6 11 13 3 58 PM ...
Page 313: ...307 51638_00048010155_C indb 307 6 11 13 3 58 PM ...
Page 314: ...308 51638_00048010155_C indb 308 6 11 13 3 58 PM ...
Page 315: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...