84
Installation Guide
Maximum switching (breaking) power
(AC1 [resistive])
1500 VA, 250 V AC
Contact resistance
100 mΩ (at 1 A/6 V DC)
Rated insulation voltage
250 V AC
Rated impulse
withstand
voltage U
imp
between coil and
contacts
4 kV 1 min
between open contacts
1 kV 1 min
Environmental specifications
Operating ambient temperature
–50 °C to 55 °C (–58 °F to 131 °F)
Operating ambient humidity
5% to 95% rh, non-condensing
Shipping/storage temperature
–40 °C to 70 °C (–40 °F to 158 °F)
Pollution degree
4 (outdoor use), 2 (internal)
Operating altitude
–610 m to 2000 m (–2001 ft to 6562 ft)
Oil temperature at valve
–20 °C to 120 °C (–4 °F to 248 °F)
Oil pressure at valve
0 to 1000 kpa/0 to 10 bar/0 to 145 psi
Mechanical specifications
Sensor head
Dimensions (L × W ×H)
392 × 264 × 158 mm (15.43 × 10.39 × 6.22 in)
Weight
8.0 kg (17.6 lb)
Interface to transformer
1.5 NpT, male thread
Enclosure
Ip67/NEMA 4X/C4
Analytical unit
Dimensions (with mount) (L × W ×H)
685 × 863 × 292 mm (26.76 × 33.97 × 11.49 in)
Weight
64.9 kg (143.3 lb)
Enclosure
Ip66/NEMA 4X
Measurement specifications
Hydrogen (H₂)
Range (ppm)
25 to 5,000
Accuracy (ppm)
±25 (or ±20%, whichever is higher)
Repeatability H₂ (ppm)
±15 (or ±10%, whichever is higher)
Response time
T90 typical 30 minutes, T90 max 60 minutes
Moisture (H₂O)
Measurement range (aw)
0 to 1 (0 to 100% rh)
Measurement accuracy (aw)
±0.02 (±2% rh)
Range (ppm)
0 to 60 @ 25 °C (77 °F) or 0 to 180 @ 55 °C (131 °F) ±3%
Temperature measurement accuracy
±0.2 °C (±0.4 °F)
Summary of Contents for CoreSense M10
Page 1: ... INSTALL ATION GUIDE CoreSense M10 Multi gas monitoring system ...
Page 18: ...Page intentionally left blank ...
Page 26: ...Page intentionally left blank ...
Page 34: ...Page intentionally left blank ...
Page 42: ...Page intentionally left blank ...
Page 62: ...Page intentionally left blank ...
Page 70: ...Page intentionally left blank ...
Page 82: ...Page intentionally left blank ...
Page 88: ...Page intentionally left blank ...
Page 93: ......
Page 94: ......
Page 95: ......