Page 24
HV/HW-CEP Series
10.
Accumulation Function
This function counts the number of times articles are weighed, calculates the total mass
value and can display the number (accumulation count) and accumulated mass value.
The accumulation function is displayed with up to 6 digits. The balance can not display 7 or
more digits, therefore the leading digits are not displayed.
Example: With 60K type, when importing the data of 17 accumulations of up to 60kg (60.000
X 17=1020.000), the balance displays this as “
020.000
”.
To use this function, set the parameters of the "Accumulation function (
5um
)" in the
function table in advance.
The accumulation count and accumulated mass value are stored in the scale even if
the power is removed.
Operation and Keys
The display of the accumulation count has a blinking
M+ without a unit.
The display of the accumulated mass value has a unit and a blinking
M+.
Press the
SET
and
M+
key to display the accumulation count and accumulated mass
value.
Press the
ZERO
key in the accumulation function (with a blinking
M+) to reset the
current function (The count and accumulated mass value become zero.)
Caution
The accumulation function is available only when weighing is performed in the
same unit.
Parameter List and Word Definition
The "Near zero" is within
±
4 digits from the zero point in the unit of kg.
The "digit" is a unit of display, and is equivalent to the minimum measurable mass.
The "zero point" is the fundamental starting point to weigh anything.
Function table
Description
A
5um 0
Accumulation function not used.
A
5um 1
The scale accumulates the data, if the
M+
key is pressed, when
the display is a positive stable value and not near zero. The next
accumulation can be performed after the display becomes near
zero or a negative value.
A
A
A
A
5um 2
The scale accumulates the data, if the
M+
key is pressed, when the
display is a stable value and not near zero. The next accumulation
can be performed after the display becomes near zero.
A
A
A
Function table
Description
A
5um 3
When the display is a positive stable value and not near zero, the
scale accumulates the data automatically. The next accumulation
can be performed after the display becomes near zero or a
negative value.
Summary of Contents for HV-15KCEP
Page 15: ...HV HW CEP Series Page 13 The label on the scale ...
Page 44: ...MEMO ...
Page 45: ...MEMO ...
Page 46: ...MEMO ...
Page 47: ......