Outline and Features
The HV-CEP/HW-CEP series are intrinsically safe electronic scales complying with
international standards such as IECEx and ATEX or FM requirements.
The HV-CEP/HW-CEP
series can be used in areas where explosive atmospheres are
present. (For details on the explosion-proof structure, refer to “6. CONFORMING
STANDARDS.”)
The HV-CEP series are platform scales with 1/3000 resolution, and have the triple
weighing range function to select the weighing range. The scale automatically
switches to small scale interval when a light sample is weighed or large scale interval
when a
heavy
sample is weighed, depending on the sample weight (multi-interval).
The HW-CEP series are platform scales with 1/10000 resolution.
The HV-CEP/HW-CEP series
use batteries.
As a power source, only the four “D”,
“LR20” or “AM1” alkaline 1.5 V batteries listed below can be used.
DURACELL MN1300, ENERGIZER E95, Panasonic LR20(XJ)
The scales use a back lit liquid crystal display to enable viewing in dim light.
The counting mode converts the total mass value (total weight) of articles to be
counted, to a count, when each article has the same mass value.
The comparator function compares the displayed weight value with the upper limit
value (HI) and the lower limit value (LO) and displays the result.
The scale has an accumulation function with a maximum of 6 digits, which has a
maximum accumulations of 999 times. (The number of times weighed and the total
mass of that can be stored in the scale.)
The following parameters are stored in the scale even if the power is removed.
Display mode (unit)
Unit mass of counting mode
Total count and total mass of accumulation function
Upper limit value and lower limit value of upper / lower comparator function
Calibration data
Parameters of the function table
Multi-interval
: This is what the minimum display is automatically switched to
depending on the sample weight.
Example) With the HV-60KCEP, weighing capacity is 60kg and minimum display is
0.005kg, 0.01kg or 0.02kg.
Multi-interval
: When exceeding the range for small, medium or large, the minimum
display is automatically switched.
Point:
When a light sample is
weighed, the minimum
display is small.
When a heavy sample is
weighed,
the
minimum
display is large.
60kg
0.005kg
15kg
30kg
0.02kg
0.01kg
Small
range
Large
range
Medium
range
Weighing capacity
Minimum
display