![A&D HV-15KCEP Instruction Manual Download Page 24](http://html.mh-extra.com/html/aandd/hv-15kcep/hv-15kcep_instruction-manual_446475024.webp)
Page 22
HV/HW-CEP Series
9.
Counting Mode
The counting mode is the function to convert the total mass value (total weight) of
articles to a count, when each article has the same mass value.
To use this function, store a unit mass in advance.
Even if the batteries are removed, the unit mass is maintained in non-volatile
memory.
9.1.
Storing a Unit Mass
Step 1 Press the
MODE
key to display the unit .
Step 2 Press the
SET
and
ENTER
key to enter the mode
that stores a unit mass.
Step 3 Press the
SET
key to select the number of samples.
The greater the quantity of samples, the greater the
accuracy of the count.
5 pieces, 10 pieces, 20 pieces, 50 pieces,
100 pieces
Step 4 Place the container on the pan.
Press the
TARE
key.
Step 5 Place the number of samples selected at step 3.
Wait for the stability mark to be displayed. Press the
ENTER
key to store. The count is displayed.
Caution
When the sample total mass value is too small and
it is not possible to calculate a unit mass, the scale
displays
lo ut
and returns to step 3. Increase the
number of samples. More than 10 digits of the total
sample mass, in the unit of kg, are required.
When the unit mass is too light to store, the scale
displays
lo ut .
In this case, the unit mass will not be
stored even if the number of samples is increased.
Pressing the
MODE
key, after
lo ut
is
displayed,
displays the next unit.
Step 6 Remove the samples and the container from the pan.
Counting mode
STABLE
STABLE
STABLE
Zeroing value
Count
Pan
A container(bowl)
Note
The pan shape depends
on the scale model.
Number of samples
STABLE