HV/HW-CEP Series
Page 33
13.2.3.
Calibration of the Zero
Point
Step 8 After 30-minute warm up, press the
CAL
switch to display
Cal
.
Press the
ENTER
switch to display
Cal 0
.
Step 9 Confirm that nothing is placed on the pan.
Wait for the stability mark to be displayed.
Press the
ENTER
key.
The scale stores the current condition as
the zero point.
Step 10 Calibration of the zero point is finished. To
exit the calibration mode, proceed to step
14.
13.2.4.
Span Calibration
Step 11 Set the value of the calibration mass
using the numerical keypad. (The initial
value depends on the scale model.)
Step 12 Place the mass on the pan which was set
at step 11. Wait for the stability mark to be
displayed. Press the
ENTER
key.
The scale then calculates the span and
stores it.
Step 13 The scale displays
Cal
a
fter displaying
end
to
finish setting.
Remove the mass from the pan.
Step 14 Press the
CAL
switch to finish calibraiton.
Weignt value
CAL switch
STABLE
Stability mark
Stability mark
CAL switch
Weighing mode
Note
The pan shape depends
on the scale model.
Nothing on
the pan
15kg
The power is automatically
turned off.
Enter using numerical
keypad
STABLE
STABLE