Page 32
HV/HW-CEP Series
13.2.
Complete Calibration Procedure
13.2.1.
Gravity Acceleration Correction
Step 1 Turn on the display.
Open the rear cover of the display unit. Locate the
CAL
switch inside.
Press the
CAL
switch to enter the calibration mode.
Then
Cal
is displayed.
Step 2 Press the
MODE
key to enter the gravity acceleration
correction mode. Press the
ENTER
key to make the
numerical value blink and proceed to input mode.
Step 3 Set your local gravity acceleration using the numerical
keypad.
Step 4 Press the
ENTER
key to store the new value.
The display returns to gravity acceleration display
after displaying
end
.
Step 5 Press the
CAL
switch again to finish gravity
acceleration correction.
13.2.2.
Preparation
Step 6 Confirm the environmental conditions as follows:
Maintain a constant temperature and stable power.
Install the scale on a solid floor where there is no draft, vibration, strong magetic
fields or direct sunlight.
Refer to "3.2. Precaution on Installing the Scale".
Step 7 Keep the display turned on for at least 30 minutes to warm up the scale.