6.1 Preparation
•
Make sure the power cord is plugged in.
•
Make sure the door contact switch is connected- or check connection to L-ORD
when using the CLASSIC 514 with L-ORD
•
Check that the Laser-Stop button is not pressed.
If it is pressed, pull it up.
•
Are third parties present who need safety glasses?
A T T E N T I O N
The CLASSIC 514
should not be operated for more than 5 hours at a time. Restart the
laser after 5 hours at the latest so that it can carry out the system check during the start
routine. This ensures that undetected errors cannot occur.
6.2 Starting the laser
The CLASSIC 514 is used together with a slit lamp. The device is connected via a power
supply to the lifting table. To start the device and the slit lamp, switch on the main switch
underneath the table. Afterwards, the device can be started using the On/Off switch on
the back of the CLASSIC 514.
When using the CLASSIC 514 with another slit lamp, lifting table respectively, switching
on may be differ from the previous description.
After the main switch at the lifting table and the device is switched on, the green status
LED on the front starts to flash. The A.R.C. Laser GmbH logo is shown on the display
and the system check starts.
Meanwhile the laser checks the
internaly safety functions. This may
take a while.
The green LED stops blinking and
lights
up
continuously
after
successfully performing the system
check.