of the joystick of the slit lamp. The article number of the foot switch is BG04008.
Accessories for the slit lamp and the table are listed in the User Manual of the slit lamp.
Laser safety goggles are not part of the basic equipment and are listed separately on the
purchase order. We recommend to order at least two laser safety goggles per device.
The article number for the goggles is AS01003.
For information about other accessories, please contact A.R.C. Laser GmbH or the
responsible sales partner.
A T T E N T I O N
Only spare parts and applicators which are approved by A.R.C. Laser GmbH are to be
used with the device. Accessories that have not been approved can significantly impair
the safety and reliability of the device.
The use of accessories, transduce and services other than those which the A.R.C.
Laser GmbH has determined or provided, may result in increased electromagnetic
interference or reduced electromagnetic immunity of the device and lead to incorrect
operation.
5.4.1 Instrument table
The instrument table on which the slit lamp is mounted is height adjustable. This allows
to adjust the optimal working height comfortably for each patient. Please note that the
table can only take an additional load of 2.0 kg. This means that no objects heavier than
2.0 kg in total may be placed on the table. Also, no persons may sit on the table or
support themselves with their entire body weight on it.
If the table is tilted, you can calibrate it. To calibrate the table, move it all the way down
and press and hold the button (down arrow) until it comes to a complete stop. After
calibration, the table is horizontal again.
5.4.2 Foot Switch (optional)
An optional footswitch is available for your CLASSIC 514. This enables the laser to be
shot by footswitch.