Technical data
8.1 CLASSIC 514
8.1.1 General
Model CLASSIC 514
Diode laser
Cooling
Air internal
Weight
3,3 kg
Dimentions
H 17 cm/B 25 cm/T 22cm with stand
Control Option
Buttons and rotary wheel
8.1.2 Laser Data
Wellenlänge
Energie
Toleranz
514 nm
50
– 1200 mW
±20% of displayed
energy according to
nach IEC 60601-2-22
Pulse length
15 ms
– 2000 ms
Frequency/ pulse pause
1-10 Hz, 100 ms
– 900 ms
Aiming Beam
635 nm red < 1mW, variable brightness
Power transmission
coupled in a slit lamp with 62.5 µm fiber
8.1.3 Electrical Connection Data
Power supply connection values 100-240 V, 50Hz/60Hz, 1.06
– 0,45 A
8.1.4 Classification
Laser class working beam
4
(Classification EN 60825-1)
Laser class Aiming Beam
2
(Classification EN 60825-1)
Classification
IIb, Rule 9
according to MDD/ MDR
Electrical protection class
II
(Classification IEC 60601-1)
Certification
CE 0123
8.2 L-ORD
8.2.1 General
Model L-ORD
external laser adaptor
Weight
580 g
Dimenions
H 14 cm / B 12.5 cm / T 19 cm
8.2.1 Laser-Data to be used with the L-ORD
Maximum laser power
1.5 W
Spot diameter
50 µm
– 500 µm
Working distance
54 mm
8.2.2 Classification
Classification
I, Rule 4
according to MDD/MDR