Laser warning labels
(attached on the side)
The device is marked with an additional label, which indicate to respect the user manual.
4.16 Operating conditions
The medical laser CLASSIC 514
is not suitable for use in connection with combustible
gas mixtures of all kinds.
The device is not approved for operation at altitudes above 2,000 m above sea level.
and only for an air pressure between 1080 hPa and 750 hPa.
The following environmental conditions must be met:
•
Ambient
temperature:
18° C to 35° C
•
Relative humidity: <75 %
•
4.17 Electromagnetic compatibility
The CLASSIC 514 laser system meets the EMC requirements according to EN 60601-1-
2. Guidelines and the manufacturer's declaration are described in Chapter 11.
A T T E N T I O N
This device should not be operated immediately adjacent to or stacked with other
devices.
This device should not be operated immediately adjacent to portable or mobile wireless
communication devices or stacked with portable or mobile wireless communication
devices.
Laser stopp label and marking of the fiber exit
(attached on the device)