User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx
24
6
Operation
This part of the manual only describes the technical application of the device, without
going into the medical application. You will receive separate application manuals from
A.R.C. Laser GmbH, if needed.
Settings and adjustments should only be made in accordance with the operating
instructions. Changes or settings that are not described in this manual can lead to
malfunctions.
The laser is in the READY mode during treatment. If you need to interrupt, the laser
must be returned to STANDBY mode. The device must always be switched off when
unattended to prevent operation by an unauthorized third party.
A T T E N T I O N
Since the aiming beam takes the same way through the laser transmission system as
the working beam, it is a good way to check the integrity of the laser transmission
system. If the aiming beam does not appear at the distal end of the laser transmission
system, its intensity is weak or if it looks diffuse, this is a possible indication of a
damaged or malfunctioning laser transmission system.