User Manual CLASSIC 514
August 2023
File: User Manual_CLASSIC 514_rev5.docx
16
It is prohibited to use the device’s socket for other usages than those detailed here.
Connector
Description
Usage
Color
Service Port
Service port, do not use
The service port may only be
used by trained A.R.C. Laser
service personal
Green
Warning light
Connection for the external
warning light of the door
Only use the A.R.C. Laser
GmbH connection for the
CLASSIC 514
Orange
Eye safety filter
Connector for the mechnical
eye safety filter
Not in use anymore, please
insert the dummy plug.
Black
Interlock
switchr
Interlock switch for the door
inter lock or fort he L-ORD
Der interlock switch can
either be used for the door
interlock or when conecting
the L-ORD with the CLASSIC
514
Red
Foot switch
Connector for the A.R.C.
Laser GmbH footswitch
Only use the A.R.C. Laser
GmbH foot switch for the
CLASSIC 514
Blue
Power supply
Connector for the medical
certified A.R.C. Laser power
supply
Only use the A.R.C. Laser
GmbH power supply for the
CLASSIC 514
Yellow
4.11 External interlock switch
A door interlock switch is required by the accident prevention regulations. The device is
equipped with an interlock connector as standard, which can be replaced by a door
interlock switch. The laser switches off when the door is opened. In addition, the laser
cannot be switched to the READY mode when the door is open. An error message
appears on the control panel. When the door is closed, the error message disappears
and the laser can be switched to READY again.
The door interlock plug should be inserted instead of the dummy plug. Only connect
plugs with 12V and 20mA.
Make sure the switch stucks firmly to prevent unexpected system interlock problems.
4.12 Safety-Shutter / Aiming beam
The CLASSIC 514
has an internal safety shutter. This shutter opens by pressing the
READY button and allows the user to release laser radiation.
Thereafter, the aiming beam is only visible when the laser is in the READY mode. This
aiming beam is a laser beam with very low power.
4.13 Manual reset
If an error occurs, the system switches into STANDBY mode. In that case, you should
switch the laser off and on again. Wait for 20 seconds when switching the device on
again otherwise the error F00f is shown and you have to restart again. The restart
should fix the error due to the automatic recalibration.
If the error still occurs, this can only be fixed by trained personnel. Please contact the
service of A.R.C. Laser GmbH in this case.
4.14 Reset by power failure
If the device is disconnected from the mains accidentally - e.g. due to a power failure - it