18
Inserting a sensor and s
tarting a session
/
English
4.12
Wait for the sensor to stabilise
Once you have inserted a new sensor, you must wait ap-
proximately 45 minutes for the sensor to stabilise before
you can calibrate the system. While the sensor is stabi-
lising, the app will display a progress bar and number
of minutes until calibration within the HOME screen and
TRANSMITTERS screen.
Once the sensor has stabilised, it is time to calibrate. If
you have allowed the app to send notifications and have
turned on the calibration alert, you will receive a notifica-
tion that it is time to calibrate. See CHAPTER 7 for details
about alerts.
The HOME screen and TRANSMITTERS screen will indica-
te that it is time to calibrate. See CHAPTER 5 for details
about calibration.
new session.
• Do not remove the transmitter while the sensor assembly
is attached to you.
• If you feel resistance when removing the sensor insertion tool
with two separate sensor assemblies, call Customer Service.
Once you have inserted a new sensor, it may take 1-3 mi-
nutes for the app to detect the sensor.
NOTE: If the edges of the adhesive patch begin to peel
up at any time during your session, you may apply the
overpatch. See SECTION 15.7 for instructions on how to
apply the overpatch.
4.11
Remove the sensor guide from
the sensor insertion tool
After you insert the sensor, you need to remove the sensor
guide from the sensor insertion tool, so you can use the
sensor insertion tool the next time you insert a sensor.
1.
Locate the release tab on the bottom
of the sensor insertion tool. Press and
hold the release tab.
2.
While pressing up on the release tab,
pull back on the sensor guide
3.
Discard the used plastic sensor guide.
Store the reusable sensor insertion
tool with the other supplies for
your GlucoMen Day CGM.
Home screen:
Transmitters screen:
Home screen:
Transmitters screen: