16
Inserting a sensor and s
tarting a session
/
English
4.9
Remove the sensor adhesive liner
Once you have cleaned the insertion site, you are ready to
remove the liner on the sensor adhesive patch. To com-
plete this step, you need the sensor insertion tool with the
sensor assembly attached.
CAUTION
The adhesive is sterile; do not touch the centre of the
adhesive material. Only touch the perimeter of the adhe-
sive when removing the liner.
1.
Hold the sensor insertion tool with
the sensor adhesive liner facing up.
Locate the tabs (A and B) on each
side of the liner.
2.
Grip the right tab and peel it back to
remove the first section of the liner.
A
B
A
B
2.
Locate the tab (B) on the sensor insertion tool at the end with the
lever (1 bump).
B
B
3.
Hold the sensor insertion tool, with
the tab (B) pointing down, perpendi-
cular to the sensor assembly.
Align the tab directly above the ope-
ning (A) on the sensor assembly.
NOTE: The sensor assembly opening
(A) is for alignment and positioning
of sensor insertion tool tab (B).
4.
Lower the sensor insertion tool, ste-
ering the tab into the opening on the
sensor assembly.
5.
Tilt the raised end of the sensor in-
sertion tool downward to meet the
sensor assembly, until you hear a
click.
6.
The sensor insertion tool and sensor guide should snap together
without a gap.
Check for a gap between the sensor insertion tool and sensor gui-
de. If there is a gap, see SECTION 15.7.