If the unit does not function properly, remove the
battery and reinsert it as described above.
Battery Charging
왘
The unit contains a powerful lithium-ion rechargeable
battery. The battery does not have any memory
effect and can therefore be recharged at any
time by placing it in the charger (see the section
describing the battery power level display on the
handpiece).
왘
The battery can also be charged in the charger
independently of the handpiece.
왘
Place the handpiece on the charger for approx.
1.5 hours before the first use, in order to charge the
new battery completely for the first time.
- The status LED showing the operating status of
the charger flashes green during the charging
operation; see below “Operating Status Display
of the Charger”.
Operating Status Display of the Charger
Power Level Display of the Handpiece
Seal
Status LED showing
operating status
Status LED
Operating status
Handpiece/battery Handpiece/battery
outside the charger in the charger
Steady green light Charger ready for Charging has
operation
been completed
Flashes green
––
Battery is being
charged
Steady red light
Charge contact
Charge contact
pins are wet
pins are wet
Flashes alternately Malfunction in the Problem during
red and green
charger
charging
Power level LED
Power level LED
Operating status
Handpiece outside Handpiece in the
the charger
charger
Steady green light Battery is charged, Not possible,
handpiece ready
handpiece is in
for operation
sleep mode
Steady red light
Warning for low
Not possible,
battery capacity,
handpiece is in
remaining capacity sleep mode
adequate typically
for 5 10-sec cycles
Flashes red
Battery fully
Problem in
discharged,
charging, battery
exposure cycle
is defective or
will be completed
cannot be charged
or, if in continuous
mode, stopped
LEDs showing
exposure times
{
START button
TIME button
Operation
Selection of Exposure Time
Exposure time options: 5, 10, 15, 20 sec, continuous
mode (120 sec), tack-cure mode.
왘
Due to the high light output of the Elipar S10, the
times of 5, 10, 15, 20 sec are equivalent to 10, 20, 30
and 40 sec using a light polymerization unit of
conventional performance (between approx. 600
and 800 mW/cm
2
for halogen technology or between
300 and 400 mW/cm
2
with LED). Thus, the normal