SEITE 158 - 148 x 210 mm - 44000185866/03 - SCHWARZ - 12-410 (sr)
158
Guidance and manufacturer´s declaration – electromagnetic immunity
The Elipar S10 Dental Light Unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the Elipar S10 Dental Light Unit should assure that it is used in such an environment.
IEC 60601
Electromagnetic environment -
Immunity test
Compliance level
test level
guidance
Electrostatic
± 6 kV contact
± 6 kV contact
Floors should be wood, concrete or
discharge (ESD)
ceramic tile. If floors are covered with
± 8 kV air
± 8 kV air
synthetic material,the relative humidity
IEC 61000-4-2
should be at least 30 %.
Electrostatic
± 2 kV for power
± 2 kV for power
Mains power quality should be that
transient / burst
supply lines
supply lines
of a typical commercial or hospital
environment.
IEC 61000-4-4
± 1 kV for input/output
± 1 kV for input/output
lines
lines
Surge
± 1 kV differential
± 1 kV differential
Mains power quality should be that
mode
mode
of a typical commercial or hospital
IEC 61000-4-5
environment.
± 2 kV common mode
± 2 kV common mode
Voltage dips, short
< 5 % U
T
< 5 % U
T
Mains power quality should be that of
interruptions and
(> 95 % dip in U
T
)
(> 95 % dip in U
T
)
a typical commercial or hospital
voltage variations
for 0,5 cycle
for 0,5 cycle
environment. If the user of the
on power supply
Elipar S10 Dental Light Unit requires
input lines
40 % U
T
40 % U
T
continued operation during power
(60 % dip in U
T
)
(60 % dip in U
T
)
mains interruptions, it is recommended
IEC 61000-4-11
for 5 cycles
for 5 cycles
that the Elipar S10 Dental Light Unit be
powered from an uninterruptible power
70 % U
T
70 % U
T
supply or a battery.
(30 % dip in U
T
)
(30 % dip in U
T
)
for 25 cycles
for 25 cycles
< 5 % U
T
< 5 % U
T
(>95 % dip in U
T
)
(> 95 % dip in U
T
)
for 5 sec
for 5 sec
Power frequency
3 A/m
3 A/m
Power frequency magnetic fields
(50/60 Hz)
should be at levels characteristic of a
magnetic field
typical location in a typical commercial
or hospital environment.
IEC 61000-4-8
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.