battery. The device is a medical electrical device in
accordance with IEC 60601-1 and is available as a
tabletop device. Wall mounting is not possible.
The light source is a high-performance light-emitting
diode (LED). In contrast to halogen lights, the emitted
light specifically covers the light wavelength between
430 and 480 nm, the relevant range e.g. for camphor
quinone-containing products.
The polymerization performance is so high that the
exposure times can be reduced by 50% in comparison
with a conventional halogen light (with light intensity
typically ranging from 600 to 800 mW/cm
2
). Selected
filling composites can be cured in as little as 5 seconds
if the light guide can be placed in close proximity to
increment. Filling composites which can be cured in
5 seconds are shown in the list included with the unit.
Settable exposure times:
• 5, 10, 15, 20 sec
• Continuous mode (120 sec)
• Tack-cure mode
The charger is equipped with an integrated light
intensity meter.
The unit is shipped with a light guide with 10 mm
diameter. It is not permissible to use the light guides of
other units.
The handpiece is equipped with a sleep mode to
minimize the unit’s energy consumption. The hand -
piece switches to sleep mode once it is placed in the
charger or if left unused for approx. 5 minutes outside
the charger.
In stand-by mode, the charger consumes maximally
0.5 W. This corresponds to the value recommended by
the European Commission pursuant to the EU Directive
2005/32/EC “Energy-using Products” for stand-by
operation (from 2008).
Do not discard these Instructions for the duration
of product use.
Fields of Application
• Polymerization of light-curing dental materials with
photo initiator for the wavelength range 430-480 nm.
- Though the majority of light-curing dental materials
are responsive in this range of wavelengths,
you may wish to contact the manufacturer of the
material in question.
Technical Data
Charger
Operating voltage:
100-127 V 50/60 Hz
230 V 50/60 Hz (see rating
plate for factory-set voltage)
Nominal consumption: 0.045 A (230 V)
0.1 A (100-127 V)
Dimensions:
Length: 170 mm
Width:
95 mm
Height: 50 mm
Weight: 650 g
Classification:
Protection class II,
Handpiece
Power supply:
lithium-ion battery,
nominal voltage 3.7 V
Utilizable wavelength
range: 430-480 nm
Wavelength peak:
455 nm +/-10 nm
Light intensity (between
400 and 515 nm):
1200 mW/cm
2
-10%/+20%
(independent of battery power
level)
Light emission area:
60-65 mm
2
(optically active)
Intermittent operation: 1 min on, 15 min off
(ambient temperature of 40 °C)
typically 7 min operating time
at ambient temperature (23 °C)
Light guide
Handpiece
Charger
Power cord
Glare shield