APPENDIX
ALARM TESTING PROCEDURE
Air-in-Line Alarm Test
1.
Open the pump door and remove the administration
set. Close the pump door.
2.
Program primary flow RATE at 500 ml/h, and
primary VTBI at 100 ml.
3.
Press
RUN/STOP
key to start the infusion.
4.
The infusion status indicator will turn blinking red,
audio alarm will sound, and the Main Display will
show ―AIR-IN-LINE‖. The infusion will be stopped.
5.
Press the
RUN/STOP
key again to acknowledge the
alarm condition. Z-800 Pump will be in PAUSE state.
―PAUSE‖ message will appear on the display screen.
6.
Clear the Air-in-Line condition. Press
RUN/STOP
key again to resume infusion.
Door Open Alarm Test
1.
Start a primary infusion. During operation, open the
pump door.
2.
The infusion status indicator will turn blinking red,
the audio alarm will sound, and the Main Display will
show ―Door Open‖. The infusion will be stopped.
3.
Press
RUN/STOP
key to acknowledge the alarm
condition. Z-800 Pump will be in PAUSE state.
―PAUSE‖ message will appear on the display screen.
4.
Close pump door. Press
RUN/STOP
key again to
resume infusion.