Z-800 Instructions for Use.
30/69
P/N 800-600-01-08 Rev 08/10
GETTING STARTED
TPN MODE INFUSION (Continued)
Programming TPN Mode Infusion
TPN infusion programming screen contains the following
display areas:
1.
Status Bar
: The Status Bar displays current operating
mode; current pump state; and volume infused.
TPN
: TPN mode operation.
SET
: Pump programming state.
VINF
: Volume Infused for the current infusion.
2.
Infusion Parameters
: TPN total VTBI and Total TIME
of the infusion.
TIME
: Total TIME of the infusion.
VTBI
: Total Volume To Be Infused.
3.
Soft Keys
: Allow user to access pump configuration and
TPN ramp programming screen.
CONFIG
: Access pump configuration.
RAMP
: Access TPN Ramp Up and Ramp Down
Time parameter programming.
Up/Down Arrows
: Highlight the selected parameter
item.
WARNING: Do not over-program VTBI. Program
the actual amount of the fluid in the IV bag.
The TPN infusion parameters may be programmed by using
the UP/DOWN Arrow keys and Data Entry keys.
Access TPN Ramp Time Parameter Programming
From TPN programming screen, press RAMP Soft-Key to
access TPN Ramp Parameter programming screen.
Programming TPN Ramp Time Parameters
The TPN Ramp Time parameters may be programmed by
using the UP/DOWN Arrow keys and Data Entry keys.