Z-800 Instructions for Use.
18/69
P/N 800-600-01-08 Rev 08/10
GETTING STARTED
CONTINUOUS MODE--PRIMARY INFUSION (Continued)
Primary Infusion Execution
During an infusion, the two Infusion Status Indicator LEDs
will illuminate in green and toggle once per second.
The primary infusion execution display contains the
following contents:
Status Bar:
The Status Bar displays the current operating
mode; current pump state; and volume infused.
CONT
: Continuous Mode Operation.
RUN
: Pump RUN state.
TL
: Time Left in HH:MM format for the current
infusion.
Infusion Parameters:
Current infusion parameters.
Pri RATE field displays primary flow rate.
Pri VTBI field counts down to show remaining VTBI.
Pri VINF field counts up to show volume infused.
3 minutes before infusion completes, a short audio prompt
tone will sound twice. Along with the audio prompt, the
screen will display ―INFUSION NEAR END‖ message.
This alert will repeat every 5 seconds until the infusion is
complete.
At completion of the infusion, an audio prompt sounds.
Along with the audio prompt, the screen will display
―INFUSION COMPLETE – KVO‖. This alert will repeat
every 5 seconds until user intervention. During KVO state,
the Flow Rate Indicator will change to display KVO flow
rate.
Clearing the Volume Infused During a Primary Infusion:
Pause the primary infusion by pressing the
RUN/STOP
key
during infusion. Exit the PAUSE screen by pressing the PROG
soft key. From the
Primary SETUP
screen, press the
CLR
soft
key. An audio/visual prompt will be presented requesting user
confirmation of clearing VINF. Press the
OK
soft key to
confirm. The VINF field will be reset to 0 ml. A confirmation
screen will be displayed briefly indicating VINF has been
cleared. From the ―Clear VINF?‖ screen, user can press the
QUIT
soft key to abort the action.
NOTE
: The VINF field will record cumulative infusion
volume infused unless cleared by pressing the
CLR
soft key or
turning the pump off/on and starting a new infusion.