Z-800 Instructions for Use.
25/69
P/N 800-600-01-08 Rev 08/10
GETTING STARTED
TIME/VOLUME Programming (Continued)
Programming in TIME/VTBI Option
Upon selection of the TIME/VTBI programming option,
the continuous mode infusion programming screen will
present Total Infusion Time and Total VTBI parameters.
The TIME/VTBI infusion programming screen contains
the following display areas:
1.
Status Bar
: The Status Bar displays current operating
mode; current pump state; and volume infused.
CONT
: Continuous Mode Operation
SET
: Programming pump state.
VINF
: Volume Infused for the current infusion.
2.
Infusion Parameters
: -- Primary Infusion Time and
Primary Volume To Be Infused.
Pri TIME
: Primary Infusion Time.
Pri VTBI
: Primary VTBI.
3.
Soft Keys
: Allows user to access pump configuration
and secondary infusion programming screen.
CONFIG
: Access pump configuration.
SEC
: Access secondary infusion programming
screen. The Secondary Infusion Parameters will be
presented as Sec TIME and Sec VTBI
Up/Down Arrows
: Select parameter items. Similar
to the RATE/VTBI programming screen, The
TIME/VTBI infusion parameters may be
programmed by using the UP/DOWN Arrow keys
and Data Entry keys.
WARNING: Do not over-program VTBI.
Program the actual amount of the fluid in the IV bag.