Z-800 Instructions for Use.
12/69
P/N 800-600-01-08 Rev 08/10
GETTING STARTED
INSTALLATION PROCEDURE
UNPACK THE PUMP
The Z-800 infusion pump is supplied complete with
A standard detachable, listed/certified IEC Hospital
Grade electrical cord.
Pole Clamp pre-mounted at 45 degree angle to the
pump.
Instruction for Use.
1.
Remove Pump module from its carton.
2.
Inspect the pump.
3.
Check to ensure pump door operates freely.
4.
Check for any loose parts.
NOTE:
If any of the following conditions are observed,
the Z-800 pump must be removed from use and inspected
by qualified personnel:
Look for any signs of physical damage from
shipping.
LED segments are not illuminated during system
power-on self-test.
Indicator lights do not illuminate.
Audio tune does not sound.
Main Display does not appear backlit, appears
irregular, or has evidence of a row of pixels not
functioning properly.
MOUNT PUMP TO IV POLE
1.
Attach the pump to an IV pole by turning the knob on
the pole clamp clockwise, or place the pump on a flat,
stable surface.