Z-800 Instructions for Use.
50/69
P/N 800-600-01-08 Rev 08/10
MAINTENANCE
CONFIGURABLE SETTINGS
Adjusting the Occlusion Pressure Threshold
Pressure Sensor detects down-stream occlusion in the
IV sets. User may specify occlusion alarm pressure
threshold in configuration screen. The occlusion alarm
threshold can be adjusted between 4 psi and 30 psi
1.
From the configuration menu, Select
8.
BIOMED OPT
.
2.
In the BIOMED OPTION menu, Select
8.1.
PRESSURE
.
3.
In the Pressure sub-menu, use
Up/Down Arrow
to adjust the occlusion alarm pressure threshold.
4.
Press
OK
soft key
to confirm the set value.
Press
QUIT
soft key
to back out of
PRESSURE sub menu. Press
QUIT
soft key
again to exit the BIOMED OPTION menu
WARNING:
The factory default settings
should be used unless qualified clinical personnel
determine that other customized settings are
appropriate and safe.