Z-800 Instructions for Use.
5/69
P/N 800-600-01-08 Rev 08/10
INTRODUCTION
FEATURES
Flow Rates
The Z-800 flow rate range is from 1 to 999 ml/h.
Free Flow Protection
Z-800 Pump has a built in free flow clamp to prevent
inadvertent free flow when the set is loaded in the pump.
WARNING:
Some Z-800 IV sets configurations do
NOT have set based free-flow protection. Make sure to
close roller clamp before removing administration set
from Z-800 pump.
Occlusion Pressure
The Z-800 Pump provides 14 levels adjustable
downstream occlusion alarm thresholds between 4 psi to
30 psi.
Secondary Infusions
Secondary infusions may be delivered at a specified
secondary delivery rate and secondary volume-to-be-
infused (VTBI) independent of the primary infusion
parameters. When secondary VTBI is greater then zero,
the pump will always execute secondary infusion first.
Automatic changeover occurs to the primary infusion
parameters when the secondary infusion is complete. A
Zyno Medical proprietary primary administration set
with a back check valve must be used.
System Configuration
The system configuration mode provides the ability for
qualified personnel to customize device settings.
Tamper Resist
The Tamper Resist feature provides a quick, one touch
lockout of the front keypad.
Volume-To-Be-Infused
(VTBI)
The volume-to-be-infused (VTBI) range is from 1 to
9999 ml in 1 ml increments.