Alarms
906-0330-01-01 Rev. D
ZOLL 330 Multifunction Aspirator Operator’s Guide
5-11
3143
Self Check Fault - Environmental Sensor
Alarm triggers when there is failure of the internal temperature sensors. When
this fault occurs, the device is no longer able to detect if it is operating outside of
the allowable temperature range. If operating inside of the temperature range
-26 to 55 °C (-13 to 120 °F) there is no effect on operation. If operating outside
this range, you should monitor the device continuously. When appropriate, you
should replace the aspirator and contact the service center for additional
information.
Mitigation/Info: Environmental Sensor: Temperature, Aspirator Operating,
Service Required, ** Contact Service Center**
3421
External Power Low / Disconnect
Alarm triggers when the external power (either AC or DC) drops below minimum
level (~11 VDC as supplied by either the AC/DC Power Supply or a direct DC
source) or power is intentionally disconnected. Since the device operates with
either external power or using its internal battery this is a low priority alarm that
clears when you press the Mute/Cancel button. Pressing the Mute/Cancel button
is the your acknowledgment that the device is operating on internal battery. If this
alarm occurs and you believe that the device is still connected to external power,
you should investigate the external power source and contact the service center
for additional information.
Mitigation/Info: Internal Battery Operation, Check Power Connection/Supply,
Monitor Battery Status, ** Contact Service Center**
3422
Battery Fault
Alarm triggers when the internal battery has been removed or communication
between the battery and CPU has failed. When external power is applied the
device is capable of operation however, loss of external power results in loss of
suction and a high priority alarm. Operating in this state should only be done
when no other alternatives are available.
Mitigation/Info: Battery Power Not Available, DO NOT Remove External Power!,
Ensure ability to manually aspirate patient **Contact Service Center**
3423
Battery Charging Fault
Alarm triggers when the battery charging circuit fails. When this alarm is active,
the battery cannot be charged. The device can only run with external power. If
power is lost, aspirator stops and a high priority alarm triggers. Operating in this
state should only be done when no other alternatives are available. Contact the
service center for additional information.
Mitigation/Info: Aspirator Operating, Power System Needs Repair, Battery
Cannot Charge, Maintain External Power, **Contact Service Center**
3430
Low Battery
Alarm triggers when the device detects that there is < 30 minutes of battery
operation remaining and no external power is connected. You should seek a
source of external power. Attaching external power the alarm clears immediately
however, alarm #3431 will occur in its place. See description of Alarm 3431
below.
Mitigation/Info: < 30 Minutes Operation, Connect External Power, Ensure ability
to manually aspirate patient **Contact Service Center**
Service Code
Alarm/Mitigation/Resolution