General Information
906-0330-01-01 Rev. D
ZOLL 330 Multifunction Aspirator Operator’s Guide
1-7
ZOLL Model 330 Multifunction Aspirator Indications for Use
Aspirator
The ZOLL Model 330 Multifunction Aspirator is a self-contained suction device designed for
removing debris from a patient's airway or respiratory system, secretions, blood, vomitus,
surgical fluids, tissue (including bone), bodily fluids and infectious materials during surgery.
Programmable intermittent gastrointestinal and abdominal wound drainage can also be
performed with the device. It is suitable for use in prehospital, hospital, mass casualty and
transport environments, including aeromedical. The aspirator is only for use by or on the order
of a physician.
Features
•
Portable (You can use the aspirator in hospital settings, aeromedical and ground transport,
mass casualty situations, and extreme environments).
•
Continuous and Intermittent modes.
•
Intuitive user interface.
•
Lightweight -- less than 9.9 lbs (4.5 k) without accessories.
•
Rechargeable battery provides over 3.5 hours of operation.
•
Adjustable vacuum - range 10 to 550 mm Hg for maximum clinical treatment flexibility
depending on operating mode selected.
•
Standby Mode (control) that minimizes power consumption between uses.
•
Smart Flow™ - allows the user to select OFF, or ON operation for reduced noise.
•
Smart Help™ - provides user instructions when an alarm is activated to guide the operator
through on-screen commands when responding to alarms.
•
Operating temperature range for extreme conditions: -26 °C to 55 °C (-14.8 to 131° F).
Warnings
General
•
The aspirator is intended for use by qualified personnel only. You should read this guide
before using the aspirator.
•
Before using the aspirator with a patient, you must test the device in its normal
configuration to ensure proper operation.
•
Do not modify this equipment without authorization of the manufacturer.
•
This operator’s guide is not meant to supersede any controlling operating procedure
regarding the safe use of the aspirator.
•
Follow all governing regulations regarding the disposal of any part of this medical device,
the handling of materials contaminated by body fluids, and shipment of the Li-ion batteries.
Aspirator
•
The aspirator can operate from its internal battery or from an external power source. When
using an external power source, position the power supply cables to avoid accidental
disconnect.
•
The use of accessories and cables other than those sold by ZOLL may result in increased
emissions or decreased immunity of this device.
•
Portable and mobile RF communication equipment may affect the performance of this
device. We describe the EMC performance for this device in the Specifications section of
this operator’s guide.
•
The aspirator may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation measures, such as re-orienting or
relocating of the device or shielding the location.