Alarms
906-0330-01-01 Rev. D
ZOLL 330 Multifunction Aspirator Operator’s Guide
5-13
3451
Battery Discharge Fault
Alarm triggers when the battery temperature reaches 75 °C (167 °F) and external
power is connected. Discharging the battery beyond this temperature could
destroy the battery and damage the device. During the alarm condition, the
device continues operation using external power. Shading the patient and
aspirator from direct sunlight might also help reduce the battery temperature.
Mitigation/Info: Battery Too Hot to Discharge, Do NOT Remove External Power!,
Ensure ability to manually aspirate patient, **Move To Cooler Location**
3452
Battery Charging Fault
Alarm triggers when the battery temperature is > 45 °C (122 °F). Charging the
battery above this temperature could destroy the battery and damage the device.
During the alarm condition the device continues to operate using external power
and if external power is lost, the device operates using internal battery power.
Shading the patient and aspirator from direct sunlight might also help reduce the
battery temperature.
Mitigation/Info: High Battery Temperature, Battery Does Not Charge When It is
Too Hot, Ensure External Power Available, Shade Patient and Aspirator, **Move
To Cooler Location**
3453
Battery Charging Fault
Alarm triggers when the battery temperature is 0 °C (32 °F). Charging the battery
below this temperature could destroy the battery and damage the device. During
the alarm condition the device continues to operate using external power and if
external power is lost the device operates using internal battery power.
Mitigation/Info: Battery Too Cold To Charge, Ensure External Power Available,
**Move to Warmer Location**
3455
Battery Fault
Alarm triggers when the device is not able to communicate with the internal
battery and external power is connected. To continue operation, the device must
remain connected external power. Use in this state should only be done if there
are no other alternatives. Contact the service center for additional information.
Mitigation/Info: Battery Communication, Do Not Remove External Power,
**Contact Service Center**
3470
Self Check Fault
Alarm triggers when the device is no longer able to communicate with the Power
Interface Module (PIM). When this fault occurs, you should monitor operation
continuously, replace the aspirator when possible. Contact the service center for
additional information.
Mitigation/Info: Power System, Power Management Fault, Monitor Power
Source, **Replace/Service Aspirator**
3480
Self Check Failure
Alarm occurs when the device detects a change in hardware (mismatched serial
numbers).
Mitigation/Info: Hardware Compatibility Failure, **Replace/Service Aspirator**
Service Code
Alarm/Mitigation/Resolution