14
e.g.[3]
The following is an example of calibrating a torque tool by experiment
1.
Check the mechanism operates
2.
QUADRATURE input
3.
1 decimal place
4.
Set scale factor to 1
5.
Zero the display
6.
Operate the tool for exactly 10 turns
7.
Record a displayed value of 456
8.
Write down numerator and denominator
Numerator
= 10
Denominator = 456
9.
Format values
Numerator = +010.0
Denominator = 0456
10.
Program scale factors
PNUM+010.0
PDEN0456
11.
Check the calibration using the
Calibration Check Procedure
on page 15.