VISUSCOUT 100-GA-GB-031219
48
Separation distances:
VISUSCOUT 100 maintains basic safety and performance when used in an
electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of VISUSCOUT 100 can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and VISUSCOUT 100 as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80
MHz
d = 1,2
√𝑃
80 MHz to 800
MHz
d = 1,2
√𝑃
800 MHz to 2,5
GHz
d = 2,3
√𝑃
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Содержание VISUSCOUT 100
Страница 1: ...VISUSCOUT 100 Mobile Retinal Camera User Manual...
Страница 2: ...VISUSCOUT 100 GA GB 031219...
Страница 55: ...VISUSCOUT 100 GA GB 031219 53 To disconnect the selected camera from VISUSCOUT 100 Client press Disconnect...
Страница 57: ...VISUSCOUT 100 GA GB 031219 55...
Страница 59: ......