VISUSCOUT 100-GA-GB-031219
47
NOTE 1:
U
T
is the AC mains voltage prior to application of the test level
NOTE 2
: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 3:
The ISM (industrial, scientific and medical) bands between 0.15 MHz and
80 MHz are 6.765–6.795 MHz; 13.553–13.567 MHz; 26.957–27.283 MHz and 40.66–
40.70 MHz.
NOTE 4:
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
A) Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
VISUSCOUT 100 is used exceeds the applicable RF compliance level above,
VISUSCOUT 100 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting
or relocating the VISUSCOUT 100
.
B) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3
V/m.
Manufacturer’s declaration – electromagnetic emissions:
VISUSCOUT 100 maintains basic safety and performance when used in the
electromagnetic environment specified below. The customer or the user of the device
should assure that it is used in such an environment.
Emissions test Compliance level Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
VISUSCOUT 100 uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
VISUSCOUT 100 is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic
emissions IEC
61000-3-2
Not applicable
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Complies