Appendix A – Electromagnetic compatibility
CAUTION
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that they
are operating normally. The device should not be operated in the vicinity of
high-frequency surgical equipment.
Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12
inches) to any part of the VISUSCOUT 100, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment
could result.
Manufacturer’s declaration – electromagnetic immunity:
VISUSCOUT 100 maintains basic safety and performance when used in the
electromagnetic environment specified below. The customer or the user of the device
should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±4 kV, ±8 kV
indirect contact
±4 kV, ±8 kV
contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for AC
power supply
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
±1 kV differential
mode
±2 kV common
mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.