VISUSCOUT 100-GA-GB-031219
27
15.
When camera is removed from the cradle it verifies image data storage erase.
It is recommended that image data storage is always erased before images are
captured of a new patient.
The camera keys function as shown in image below:
Fig. 7
Camera Keys
Table below provides explanations for the key functions:
Key
Press
Function
Explanation
Left soft key
Short
Control Fixation
target level and
selection
Fixation target is ON by default and
can be turned OFF by pressing left
soft key. Fixation target light has
two levels: Low and High. If the
patient cannot see the light on low
level turn it up to high.
Long
Power On / Off
Camera is powered on and off by
pressing the Left soft key for longer
than 2s.
Very
long
Reset
Camera reset is done by pressing
the Left soft key for 13 seconds
Right soft key
Short
Manual / Auto /
AF assist
Switch between focus modes by
pressing Right soft key. Manual
focus is on by default. Autofocus
range is from -15 to +10 diopters
and manual focus range is -20 to
+20 diopters.
Long
Open Menu
Enter camera menu by pressing
Right soft key for longer than 1s.