Installation
000000-1272-175-GA-GB-060613
47
•
For combined operation the VISULAS YAG III must be connected by a
special combi cable to the VISULAS 532s. For this purpose insert the end
of the cable with the single plug into the 532s socket on the YAG III laser
console (4, Fig. 8). Insert the larger of the two plugs at the other end of
the combi cable into the socket in the slit lamp (1, Fig.
) on the
VISULAS 532s, the smaller plug into the Combi socket of the
VISULAS YAG III YAG (3, Fig.
•
Connect the remaining mains and instrument cables according to the user
manuals for the individual instruments.
Cables and the corresponding sockets and ports are color-coded
for easy identification. All connectors and sockets are marked with
a red dot. These dots must be vertically aligned when they are
connected. All sockets are labeled.
CAUTION - PROPERTY DAMAGE
Connect only cables and sockets that are the same color and size. Never use
force.
To release the cable, grasp the connector by the sleeve and pull it
directly backwards. Pulling on the sleeve releases the lock
mechanism on the connector.
•
Install the door interlock on the VISULAS 532s (5, Fig.
). If you do not
wish to install an interlock, connect the yellow coding plug provided in
place of the interlock (yellow connector, see VISULAS 532s user manual).
•
The power cable for the instrument table should be connected to a fixed
power socket.
•
Connect the VISULAS YAG III to one of the sockets on the instrument
table (see separate user manual for the instrument table). The
VISULAS 532s must be plugged into a separate fixed earthed socket.
•
Remove the protective cap from the optical fiber of the applicator.
•
Insert the optical fiber carefully into the socket for the optical fiber on the
VISULAS 532s console (see VISULAS 532s user manual) and screw in
firmly.
•
Fold out the control panel foot, unwind as much of the control panel
cable as required and connect it to the front of the instrument (see
VISULAS 532s user manual).
•
Adjust the foot of the control panel to the required angle and place the
control panel in the location required.