Maintenance and care
000000-1272-175-GA-GB-060613
96
Replace the halogen bulb (5, Fig. 51) as illustrated in Fig. 51.
•
The lamp housing can be removed by pressing down and simultaneously
pulling the lug (1, Fig. 51).
•
To unlock the lamp holder, turn the lever (8, Fig. 51).
•
Pull the socket plug (4, Fig. 51) from the bulb (5, Fig. 51).
•
Then remove the defective bulb (5, Fig. 51). Push a new bulb into the
opening, paying attention to the locating pins (7, Fig. 51). Do not touch
the bulb with the bare fingers.
•
Push the socket plug (4, Fig. 51) onto the pins of the bulb (5, Fig. 51).
•
Lock the lamp holder again with the lever (8, Fig. 51).
•
Insert the cover with holder (3, Fig. 51) into the slot (6, Fig. 51) of the
lamp housing and press it in until the retainers (2, Fig. 51) engage.