Daily use
000000-1272-175-GA-GB-060613
57
Function test
When the instrument is switched on using the key switch (1, Fig. 7) on the
VISULAS YAG III laser console, a system test will be automatically performed.
During the system test the following actions, tests and initialization routines
will be executed:
•
Switch-on and initialization of the system
•
Start of system monitoring
•
Test and subsequent activation of watch dog circuit
•
Test of energy power down
•
Execution of internal laser shots and determination of energy values for
all pulse modes.
When the system test has been successfully completed, the program
switches automatically to the treatment mode.