General warning and safety instructions
000000-1272-175-GA-GB-060613
21
WARNING - RISK OF OPTICAL RADIATION
Because prolonged intense light exposure can damage the retina, the use of
LSL 532 or LIO Trion/532s for ocular examination should not be
unnecessarily prolonged, and the brightness settings should not exceed
what is needed to provide clear visualization of the target structures.
The light emitted from this instrument is potentially hazardous. The longer
the duration of exposure and the greater the number of pulses, the greater
the risk of ocular damage. Exposure to light from these instruments when
operated at maximum output will exceed the safety guideline after 3.8 min.
If you have any questions concerning the permissible prolonged exposure
time with other filter positions, please contact Carl Zeiss Meditec customer
service.
A 90 D contact glass may not be used.
Avoid looking directly into the light emitting prism.
Take particular care, when examining infants and aphakic patients.
Never look into the sun through the binocular tube and eyepieces.