Description of the device
000000-1272-175-GA-GB-060613
31
The laser console (2, Fig. 7) is the central power supply and control unit for
the VISULAS YAG III and the LSL YAG III laser slit lamp.
The control panel (5, Fig. 7), the LASER STOP button (4, Fig. 7) and the
combined control knob/button (3, Fig. 7) are located on the front side of the
control unit. The power supply and control connectors required for
operation of the VISULAS YAG III are located on the rear panel of the laser
console (Fig. 8).
The VISULAS YAG III is menu operated via the icons on the control panel (5,
Fig. 7) and the combined control knob/button (3, Fig. 7). The control
knob/button is used to adjust the system parameters in the individual
operating modes by turning the knob. Pressing the button selects the
Energy parameter in all modes.
Numerous functions and values of the system are electronically monitored. If
any of these values are outside the permitted range, the system prevents the
laser from being fired. Laser triggering, for example, is automatically locked,
when the metered energy of the therapy beam is outside the pre-defined
range.
Using the key switch (1, Fig. 7) you can prevent that unauthorized persons
switch on the laser.
The user must remove the key after the treatment.
The LASER STOP button (4, Fig. 7) is intended for control of unforeseeable
hazardous situations. When this button is pressed, the system goes into
STANDBY mode and all functions on the control unit are locked. To resume
operation, press the LASER STOP button again.
The possibility of triggering the laser can be made dependent on whether a
remote contact is open or closed, e. g. on a door. A Carl Zeiss Meditec
authorized specialist will be pleased to assist, if you wish to connect the laser
to a remote interlock.
If the instrument is not connected to a door interlock contact, the dummy
connector which bridges the pins is inserted in the interlock socket. The
instrument is delivered with the dummy plug inserted.