![background image](http://html1.mh-extra.com/html/zeiss/visulas-yag-iii/visulas-yag-iii_user-manual_3369486071.webp)
Operation of the device
000000-1272-175-GA-GB-060613
65
The buttons in the Treatment mode have the following functions:
Displays the selected focus shift mode
ANT - anterior focus shift
POST - posterior focus shift
0 - Zero focus shift
Refer to section
on page 88.
Laser emission display
This field lights up in yellow when the instrument is in
treatment READY mode and the aiming beam is
switched on. This display lights up in red when the laser
is triggered.
Selection and display of pulse mode (burst)
(shown here: single pulse mode activated)
Single pulse
Double pulse
Triple pulse
Selection and display of energy levels
Energy levels
When this field is activated, the energy can be set using
the control knob. There are 22 energy levels.
In single-pulse mode, the lowest level corresponds to
about 0.3 mJ, the highest to about 10 mJ.
The energy display indicates the average energy of the
last five shots, calculated for the currently selected
energy level. The output energy of the Nd:YAG laser may
vary slightly. You should therefore check the displayed
energy value regularly.
Energy and pulse counter
Display field
Displays the sum of the emitted pulses and the
cumulative energy.
Reset button
Touching this button resets the sum of emitted pulses
and the cumulative energy to zero.
Information mode
Switch to information mode
Touching this button switches to Information mode. See
page 66.