Installation
000000-1272-175-GA-GB-060613
50
1 Connection socket for slit lamp or physician’s safety filter
2
Connector for combined operation with the VISULAS YAG III
3 Door interlock connector
4 "Foot 1" - Connection for foot switch (standard)
5 "Foot 2" - Connection for foot switch with power control (optional)
Fig. 19
Connections on the VISULAS Trion laser console
Cables and the corresponding sockets and ports are color-coded
for easy identification. All connectors and sockets are marked with
a red dot. These dots must be vertically aligned when they are
connected. All sockets are labeled.
CAUTION - PROPERTY DAMAGE
Connect only cables and sockets that are the same color and size. Never use
force.
To release the cable, grasp the connector by the sleeve and pull it
directly backwards. Pulling on the sleeve releases the lock
mechanism on the connector.
•
Install the door interlock on the VISULAS Trion (3, Fig. 19). If you do not
wish to install an interlock, connect the yellow coding plug provided in
place of the interlock (yellow connector, see VISULAS Trion user manual).
•
The power cable for the instrument table should be connected to a fixed
power socket.
•
Connect the VISULAS YAG III to one of the sockets on the instrument
table (see separate user manual for the instrument table). The
VISULAS Trion must be plugged into a separate earthed mains socket.
•
Remove the protective cap from the optical fiber of the applicator.