CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
ApoTome.2
Cleaning the device
Carl Zeiss
05/2012
423667-7144-001
27
5
CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
5.1
Cleaning the device
Before starting cleaning works, switch off the device and disconnect it from the mains.
Ensure that no moisture enters the device.
The care of the ApoTome.2 is limited to the works described below:
•
After every use, switch off the device and cover it with the protective cover (protection from dust and
moisture). If necessary, store the ApoTome.2 slider in the provided case.
•
Do not install the device in a damp room; the maximum permissible humidity is 75 %.
•
Remove dust and loose contamination from visible optical surfaces with a fine brush, rubber blower,
cotton bud, optics-cleaning tissue or a dust-free cotton cloth without exerting pressure.
•
Wipe off water-soluble contamination (coffee, cola, etc.) after breathing by using a dust-free cotton
cloth or a moistened cloth. The water used may contain a dash of a mild detergent.
•
Wipe off stronger oily or fatty contamination (immersion oils, fingerprints) with a cotton bud, lens
cleaning paper or a dust-free cotton cloth using optics cleaning agent L.
This cleaning agent is made of 90 percent by volume gasoline and 10 percent by volume isopropanol
(IPA). Its individual constituents are also known under the following synonyms:
gasoline:
surgical spirits, petroleum ether
isopropanol:
2-propanol,
dimethyl carbinol,
2-hydroxypropane
Clean the optical surfaces parallel to the alignment of the grid exerting a slight pressure onto the surface.
Clean the transmission grid with extreme care to avoid the damage of the grid structures!
For the use of the ApoTome.2 in humid climatic zones, observe the following instructions:
•
Store the device in a bright, dry and well ventilated room; humidity < 75 %. Store components and
accessories that are especially susceptible to humidity, such as objectives and eyepieces, in drying
chambers.
Precision-mechanical optical instruments are always endangered by mold attack under the following
conditions:
•
relative humidity of > 75% longer than three days at temperatures of +15
°
C to +35
°
C,
•
installation in dark rooms without movement of air,
•
dust deposits and fingerprints on optical surfaces.