CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
ApoTome.2
Troubleshooting
Carl Zeiss
05/2012
423667-7144-001
29
5.3
Troubleshooting
Problem
Cause
Troubleshooting
Device does not respond to any
control command
no current
Actuate power switch, connect
power cable to power outlet.
no connection to the PC
Establish CAN BUS cable
connection to the stand or the
USB cable connection to the PC.
no connection between slider and
control box
Plug the cable of the slider into
the respective connector on the
control box.
device defect
Call Technical Service.
No transmission grid visible in
eyepiece although the slider is
in the ApoTome position (click
position 2)
transmission grid not focused
To obtain an artifact-free result
image, the transmission grid must
be imaged with a high quality in
the specimen plane. For this
purpose, calibrate the system as
described in the Operating
Manual of the imaging software
AxioVision 4.8.2 / ZEN (blue
edition).
5.4
Wear parts and tools
Designation
Catalogue no.
Purpose
Fuse type
T 2 A/H 250 V
000000-0149-688
Function of control box
Reflector module
calibration "ApoTome" ACR P&C
424930-9902-000
Calibration of phase position of scanner
Mirror test specimen
000000-1182-440
Calibration of phase position of scanner
Fluorescence test specimen
000000-1213-943
Calibration of the focus position of the
transmission grid for various wavelengths
Cover plate
423666-0102-000
Replacement of the original cover plate
on Axio Imager