INSTRUMENT DESCRIPTION
ApoTome.2
Technical data
Carl Zeiss
05/2012
423667-7144-001
21
2.7
Technical data
Dimensions (width x depth x height)
ApoTome.2 slider for Axio Imager ...................................................... approx. 278 mm x 90 mm x 76 mm
ApoTome.2 slider for Axio Observer/Axiovert 200 ............................... approx. 295 mm x 90 mm x 78 mm
ApoTome.2 slider for Axio Zoom.V16 ................................................. approx. 278 mm x 90 mm x 76 mm
ApoTome.2 control box ................................................................... approx. 255 mm x 220 mm x 96 mm
Weight
ApoTome.2 slider ............................................................................................................. approx. 1.1 kg
ApoTome.2 control box .................................................................................................... approx. 2.9 kg
Ambient conditions
Transport (in packing)
Permissible ambient temperature ....................................................................................... -40 to +70 °C
Storage
Permissible ambient temperature ........................................................................................ +5 to +40 °C
Permissible relative humidity (without condensation) ................................................ max. 75 % at +35 °C
Operation
Permissible ambient temperature ........................................................................................ +5 to +40 °C
Permissible relative humidity ................................................................................... max. 75 % at +35 °C
Air pressure .............................................................................................................800 hPa to 1060 hPa
Operating altitude .............................................................................................................. max. 2000 m
Pollution degree ................................................................................................................................... 2
Functional data
Area of use ..........................................................................................................................closed rooms
Protection Class ......................................................................................................................................I
Degree of protection ....................................................................................................................... IP 20
Electrical safety ....................................................................... as per DIN EN 61010-1 (IEC 1010-1) and in
compliance with CSA and UL regulations
Overvoltage category ............................................................................................................................. II
Radio interference suppression ............................................................................ as per EN 55011 Class B
Noise immunity .................................................................................................... as per DIN EN 61326-1
Line voltage ............................................................................................................. 100 to 240 V ±10 %
Line voltage adjustment is not necessary!
Line frequency ...................................................................................................................... 50 to 60 Hz
Power consumption of ApoTome.2 ........................................................................................ max. 50 VA